H-5223 – Xylem H-5223 User Manual

Page 16

Advertising
background image

2.2.2 Com Port RS-232

This is the main RS-232 port for communicating to the H-5223. The RS-232 port is used to
connect to a PC, PDA, modem, or other type of serial equipment for standard serial
communications. This port is configured as a DCE type of device. This means it will plug
directly into a PC (a DTE type device), but will require a NULL modem adaptor for connecting
to most PDA’s or a modem (a DCE type device). The NULL modem cable crosses the
communication lines allowing two similar devices to communicate. Figure 2-2 shows the pin
assignments for this port.

SERIAL PORT PIN-OUT

PIN

DIRECTION

NAME (DCE)

1

No Connect

Data Carrier Detect
(DCD)

2

Output

Receive Data

(RD)

3

Input

Transmit Data

(TXD)

4

Input

Data Terminal Ready

(DTR)

5

Ground

(GND)

6

Output

Data Set Ready

(DSR)

7

Input

Request To Send

(RTS)

8

Output

Clear To Send

(CTS)

9

No Connect

Ring Indicator

(RI)

Figure 2-2 Serial Port Pin Assignments

2.2.3 Analog Input Section

The first seven connections on the terminal block are used for analog input functions. This
includes four analog inputs, two analog grounds and one +5.00 volt excitation connection.

2-2 Hardware Options and Installation

H-5223

Advertising