Xylem WQ-FDO Sensor and GL500-7-2 Data Logger User Manual

Page 23

Advertising
background image

Global Water

800-876-1172

globalw.com

- 23 -

Disp Low: This field sets the lower limit of the bar graph displayed in the main

screen and provides a way of scaling the graph beyond the lower limit used to

calibrate the logger. In this case the lower calibration was zero.

Decimal Places: From the pull-down menu, select the

number of decimal places to the right of the decimal

point that should appear in the historical data and in the

numerical data field at the bottom of the bar graph.

Calibrate: The “Calibrate” button provides a simple 4-step process for calibrating

the GL500 Data Logger to a particular sensor. Click the Calibrate button and follow

the menu instructions.

Click on the High EU field and enter the

high engineering unit that will be used

to calibrate the sensor, not the highest

reading the sensor can output. In the

example of pH, enter 10 which is the

highest calibration solution to be used.

Click Next.

Place the sensor in the condition that will

cause it to output the level corresponding

to the high EU entered in the previous menu.

Wait for the data reading to stabilize.

Click Next.

Advertising