Xylem WQ-FDO Sensor and GL500-7-2 Data Logger User Manual

Page 28

Advertising
background image

Global Water

800-876-1172

globalw.com

- 28 -

In the DO Setup tab the input channel number and units are defined; however, this is

only for the purpose of telling the software which channel data to use for

calculations. It is in the Analog setup tab that the calibration of each sensor is

performed, see section XI. If the DO sensor is physically connected to channel 1 of

the data logger; it must be defined in the DO tab as channel 1, be properly calibrated

on channel 1 in the analog setup tab, and the units in each tab must match. A

warning message will appear when the units between these two tabs do not match.

This is an example of a message warning

that the units for the WQ-FDO sensor on

channel 1 do not match in the Analog and

DO tabs. The units must match and be

spelled the same.

Below is an example of the Analog Setup screen for the FDO sensor system. Note

that the display channels for each sensor and calculated results match between this

screen and the DO Setup tab previously shown. Also that the Enable checkboxes

are checked to allow the display and download of the data. The WQ-FDO sensor is

required for this software to operate, both the partial pressure and temperature

outputs of the sensor must be connected to the logger. They need not be enabled

for display, but they must be connected. If the optional conductivity and barometric

pressure sensors are not connected and a manual entry is used, the unused analog

channels can be connected to any other sensor. If any of the calculated output

values of saturation, concentration and salinity are not wanted for display and

download, they do not need to be enabled and the sensor channel can be used for

any other type of sensor. If they are displayed, the analog channel they are

assigned to is unavailable, even

though there is no sensor

connected to it.

Calibration is required on all channels

that have a sensor connected.

Calculated channels of concentration,

saturation and salinity do not require

calibration values of Low and High

EU and Raw Data. Low and High

Display values for these calculated

numbers scale the display only.