partial pressure of the oxygen in the water. The sensor should read

about this pressure. For example, at sea level the average

atmospheric pressure is 1013mbar. The partial pressure of oxygen in

water exposed to air for a period of time will be about 0.21 x 1013mbar

or 212mbar. Because factors like atmospheric pressure and

temperature are constantly changing, this is not a recommended way

of calibration. This should be done for performance checks only.

e. Besides oxygen partial pressure, temperature is also important in

calculating oxygen levels. The WQ-FDO sensor has a temperature

output than can be checked. Compare the temperature reading of the

sensor in water to that of another measurement device. A stable

temperature can also be made with ice water. A mixture of water with

lots of ice being stirred constantly can achieve a stable ice water

temperature of 32ºF (or 0º C) +.2F.

f. Check the logger and other sensor calibrations and recalibrate if

necessary, saving the setup file first as described in section IX.

Because dissolved oxygen is dependent on temperature, pressure and

conductivity; it is important to have accurate data for the software to

make these calculations. Save the new setup file under a different

name each time the logger is recalibrated.

Issue: Cannot communicate with the logger.

a. Verify that sufficient power is applied and wired correctly.

b. Check the communication cable.

c. If the connection is through the USB port, use the Windows Device

Manager to confirm that the driver is loaded and the correct virtual

COM port is selected in the connection menu of the Global Logger

software, as described in sections III and IV.

d. Check the Device Manager in your operating system and confirm that

the USB COM port is not being used by another device.

e. Try reducing the baud rate in the connection screen of the Global

Logger software.

f. Try switching from RS-232 to USB or USB to RS-232.