Cond 3210
Operation
33
ba75794e01
01/2009
Measurement /
Temp. comp. (TC) /
Linear coeff.
0.000 ...
3.000 %/K
Coefficient of the linear
temperature compensation.
This setting is only available
when the linear temperature
compensation is set.
Measurement /
Temp. comp. (TC) /
Reference temp.
20 °C
25 °C
Reference temperature
This setting is only available
for the measured
parameters,
ӕ
und
ρ
.
Measurement /TDS
factor
0.40 ... 1.00
Factor for TDS value
Measurement /
Stability control
On /Off
Switches on or off the
automatic stability control
during measurement (see
section 4.3.3)
Measurement /
Temperature unit
°C
°F
Temperature unit,
degrees Celsius or degrees
Fahrenheit.
All temperatures are
displayed with the selected
unit.
Reset
-
Resets all sensor settings to
the delivery condition (see
section 4.6.1).
Menu item
Possible
setting
Description