Xylem COND3210 HANDHELD CONDUCTIVITY METERS User Manual

Page 49

Advertising
background image

Cond 3210

Technical data

49

ba75794e01

01/2009

Note
The accuracy values specified here apply exclusively to the meter. The
accuracy of the measuring cell has also to be taken into account.

Accuracy (± 1 digit)

Parameter

Accuracy

Temperature of the test
sample

ӕ and

ρ

/ temperature compensation

None (Off)

± 0.5 %

Nonlinear (nLF)

± 0.5 %

0 °C ... + 35 °C
according to EN 27 888

± 0.5 %

+ 35 °C ... + 50 °C
enhanced nLF function

Linear (lin)

± 0.5 %

+ 10 °C ... + 75 °C

SAL / range

0.0 ... 42.0

± 0.1

+ 5 °C ... + 25 °C

± 0.2

+ 25 °C ... + 30 °C

TDS [mg/l]

± 0.5 %

T [°C] / temperature sensor

NTC 30

± 0.2

PT 1000

± 0.3

Advertising