Operation – Xylem 407119A HOT WIRE ANEMOMETERS User Manual

Page 5

Advertising
background image

Model 407119A Version 4.1 October 2005

5

Operation

Initialization and Zero procedures (please follow these steps before use)

Important Note

: The meter’s probe locking switch must

be in the locked position (with the probe inserted) for
the meter to turn on and operate correctly. Refer to the
accompanying diagram or the diagram at the top of the
meter for switch orientation.
For the best accuracy and for proper operation, the
following steps must be performed.

1. Connect the sensor to the input jack on top of the

meter by first unlocking the sensor jack and then
inserting the sensor plug. Lock the sensor jack after
the probe is inserted.

2. Press the POWER button to turn on the meter. The meter will perform a self-test during

which the display will count down from 9999 to 0000.

3. Select the air velocity function using the FUNCTION button. The LCD will display the

most recently selected unit of measure for air velocity.

4. To change the temperature units of measure (C/F), use the advanced programming

mode discussed later in this manual.

5. Momentary presses of the UNIT button changes the air velocity units. The LCD will

reflect the selection.

6. Place the sensor cover in the up (ZERO) position (see diagram below).
7. Open the telescoping sensing antenna to a convenient length.
8. Place the sensor in the area to be measured and allow a short time for it to stabilize to

the ambient temperature.

9. To Zero the meter display, press and hold the ZERO button until the second beep.

Warning!

Do not touch the air velocity or temperature thermistors inside the sensor head

while moving the sensor cover.

TOP VIEW OF METER

LOCKED Position

UNLOCKED Position