Model 407119A Version 4.1 October 2005

9

PC Interface

The 407119A is equipped with a 3.5mm jack (side compartment) for connection to a PC for
data acquisition purposes. The meter’s data output is a 16 bit data stream. To obtain PC
interface cabling and Windows

TM

data acquisition software, contact Extech Instruments.

Instructions for use are provided with the software. The PC interface cable schematic, data
stream protocol, and RS-232 communication settings are provided below.

PC Interface Cable schematic











Protocol for 16-bit data stream

D0

End Word = ‘0D’

D1 & D8

Displayed reading; D1 = LSD; D8 = MSD. For example,
For a displayed reading of 1234, D8 through D1 is 00001234

D9

Decimal point (right to left):
0 = no decimal; 1 = 1 place; 2 = 2 places; 3 = 3 places

D10

Polarity: 0 = positive; 1 = negative

D11 & D12

Displayed unit of measure: 01 =

o

C; 02 =

o

F; 08 = m/s; 09 = knots;

10 = km/hr; 11 = ft/min; 12 = MPH; 84 = CMM; 85 = CFM;
0A = Square meter (area); 0B = Square feet (area)

D13

1 = Upper LCD display digits; 2 = Lower LCD display digits

D14

Always ‘4’

D15

Start word ‘02’

RS-232 Settings

Baud rate: 9600

Parity: No parity

Data bits: 8

Stop bits: 1

Meter

3.5mm jack

Center pin.............................Pin 4

Ground / Shield.....................Pin 2

PC

DB-9 pin

Pin 5

2.2K

Resistor