Messages:
Message Significance
REC-ERROR: Please perform a power-on reset !
A problem occurred during the recording.
The
battery must be removed and reinstalled.
BAT-ERROR: Please replace battery !
The battery must be replaced
ACK-ERROR: Please perform a power-on reset ! There was a problem in saving during the
recording. The battery must be removed and
reinstalled.
Device has changed, please read configuration first! The connected device was swapped out.
It must be reinitialized with Read Configuration
No communication! Please check connection and
port-no.
No communication with the connected device.
The serial ports at the device and PC need to be
checked.
State: Record active
Device is in recording mode
State: Record prepared Device is being prepared for recording mode
State: Record stopped Recording mode is stopped
Memory full! The memory is completely full
(only possible, if "Endless recording (circular