Wear parts and accessories – YSI TruLine ORP 50 User Manual
Page 11
T 814241
04/2014
11
TruLine
Wear parts and accessories
Oxidizing
pretreatment
Reducing
pretreatment
Immerse the platinum electrode for two to three
days in a sulfuric acid chlorine solution. Chlorine
powder for producing the solution is included in the
SORT/RH reagent set.
Note: The diaphragm must not be immersed in the
chlorine solution!
When the electrode is ready for the test immerse it
in the RH 28 ORP buffer solution and wait for a sta-
ble measured value.
Wear parts and accessories
Description
Model
Order no.
Reference electrolyte solution 3 mol/l KCl,
Ag
+
-free (250 ml)
KCl-250
109 705Y
Reagent set for regenerating ORP platinum
electrodes, consisting of 10 g activation
powder and 30 g chlorine powder
SORT/RH
109 730Y
Zobell ORP buffer solution (125 ml)
3682
061320