Commissioning, measuring, testing

 Make sure the immersion depth is correct. The junction must be complete-

ly submersed in the solution. The junction is in the area of the bottom end
of the shaft (see arrow).

The level of the reference electrolyte must be at least 2 cm above the level
of the solution.

Conversion to the

standard hydrogen
electrode potential

U

H

= U

Meas

+ U

Ref

where:

U

H

= ORP voltage, relative to the

standard hydrogen electrode

U

Meas

= measured ORP voltage

U

Ref

= voltage of the reference system relative to the

standard hydrogen electrode

U

Ref

is temperature dependent and can be taken from the following table

(also refer to DIN 38404-6):

T (°C)

U

Ref

[mV]

TruLine ORP 50

T (°C)

U

Ref

[mV]

TruLine ORP 50

0

+224

35

+200

5

+221

40

+196

10

+217

45

+192

15

+214

50

+188

20

+211

55

+184

25

+207

60

+180

30

+203