Commissioning, measuring, testing
Make sure the immersion depth is correct. The junction must be complete-
ly submersed in the solution. The junction is in the area of the bottom end
of the shaft (see arrow).
The level of the reference electrolyte must be at least 2 cm above the level
of the solution.
Conversion to the
standard hydrogen
electrode potential
U
H
= U
Meas
+ U
Ref
where:
U
H
= ORP voltage, relative to the
standard hydrogen electrode
U
Meas
= measured ORP voltage
U
Ref
= voltage of the reference system relative to the
standard hydrogen electrode
U
Ref
is temperature dependent and can be taken from the following table
(also refer to DIN 38404-6):
T (°C)
U
Ref
[mV]
TruLine ORP 50
T (°C)
U
Ref
[mV]
TruLine ORP 50
0
+224
35
+200
5
+221
40
+196
10
+217
45
+192
15
+214
50
+188
20
+211
55
+184
25
+207
60
+180
30
+203