Maintenance and cleaning

TruLine

ing, the responding behavior becomes slower and the electrode slope and
asymmetry change. Moreover, extreme operating conditions can consider-
ably shorten the lifetime of the combination electrode. These are:

 Strong acids or lyes, hydrofluoric acid, organic solvents, oils, fats, bro-

mides, sulfides, iodides, proteins

 High temperatures

 High changes in pH and temperature.

The warranty does not cover failure caused by measuring conditions and
mechanical damage.

Refilling the

reference

electrolyte

Maintenance and cleaning

During operation, a small amount of reference electrolyte leaks through the
junction from the combination electrode into the test sample. If the level of ref-
erence electrolyte becomes too low with time, refill it through the refilling
opening. Refilling is very easy with the enclosed dropping bottle. Proceed as
follows:

 Cut off the tip of the dropping bottle at a right angle until the opening in the

tip can be seen

 Open the refilling opening of the combination electrode

 Press the tip of the dropping bottle into the refilling opening while turning

it slightly

 Pump the reference electrolyte in the shaft using the dropping bottle

 Pull the dropping bottle out of the refilling opening while turning it slightly

if necessary.