Wear parts and accessories – YSI TruLine 15 User Manual

Page 9

Advertising
background image

T 814239

04/2014

9

TruLine

Wear parts and accessories

Cleaning

Remove water-soluble contamination by rinsing with deionized water. Re-

move other contamination as follows:

Contamination

Cleaning procedure

Fat and oil

Rinse with water containing household
washing-up liquid

Lime and hydroxide deposits Rinse with citric acid (10 % by weight)

Note
Hydrofluoric acid, hot phosphoric acid and strong alkaline solutions destroy
the glass membrane.

After cleaning

Rinse the combination electrode with deionized water and condition it in

reference electrolyte solution for at least 1 hour. Then recalibrate the combi-
nation electrode.

Wear parts and accessories

Description

Model

Order no.

Reference electrolyte solution 250 mL
(KCl 3 mol/L, Ag

+

free)

KCl-250

109 705Y

Advertising
This manual is related to the following products: