2-4 Hardware Options and Installation

XL™ Series

Figure 2-2 Conoflow Gas Purge Stream Gauge Installation

DIFFERENTIAL
REGULATOR

PRESSURE
RELIEF VALVE

POWER

30 BUBBLES
PER MINUTE

SITE-FEED

GAS PURGE
LINE ORIFICE

GAS PRESSURE
TO MANOMETER

H-350XL™

20 - 30 PSI

MOUNT
ON WALL

FEEDBACK TUBE

NITROGEN
SUPPLY
CYLINDER

REGULATOR

TO RIVER

2.2.3.1 Sample Installation for the Conoflow System (H-350XL only)

Figure 2-2 shows a typical H-350XL™ installation for water depth measurement using the
Conoflow gas purge system. To install the H-350XL™, secure it to the wall of the instrument
shelter or bench top using the provided mounting hardware. This will prevent it from moving or
shifting and pulling on the wires and tubing connected to other equipment. The H-350XL™
should be mounted so moisture and dust will not settle on the main I/O panel. Normally
vertically is the best with the main I/O panel facing down and the display keypad facing out.