Overview

The Global Water WQ-COND Conductivity Sensor is suitable for measuring

conductivity in a wide variety of applications including laboratories, streams,

rivers, and groundwater. It’s small size and rugged housing make it useful for

hand held measurements or permanent installation.

The sensor uses a 4-electrode measuring technique that provides accurate

readings over a wide range of conductivities and temperatures. Because the

conductivity of ionic solutions increases with increasing temperature, a

temperature sensor is also incorporated and is used to provide automatic

temperature compensation of 2%/ºC normalized to 25ºC. An in-line interface

module converts the digital conductivity and temperature data into two

separate 4-20mA signals for monitoring with data loggers and PLC devices.

The standard sensor comes with one meter of cable between the sensor and

interface module, and 25 feet of marine grade cable for connecting to

recording devices. Additional cable can extend the length up to 1000 feet.

Available conductivity ranges are 0-200uS/cm, 200-2000uS/cm, 2-20mS/cm,

20-200mS/cm and 200-2000mS/cm. The temperature output has a

measurement range of -5ºC to +70ºC.