III. Maintenance

Clean the sensor before use, especially when measuring low conductivity

solutions. The power should be disconnected before cleaning.

For cleaning mineral deposits, immerse the sensor in acetic acid for 5

minutes. Other types of contamination such dirt and oil can be cleaned using

mild soapy water.

Normally the conductivity sensor does not age and will not require periodic

calibration. Measuring strong acids, bases or organic solvents can

considerably reduce the sensor’s lifetime or cause permanent damage. Long

term exposure to very high temperatures may also damage it. These types of

damage are not covered under warrantee.

IV.

Trouble Shooting

There are no user serviceable parts inside this sensor

Issue: Sensor reading incorrectly

a. Check all connections

b. Inspect the cable for damage

c. Check that the supply voltage is within specifications

d. Make sure the conductivity and temperature of the solution is within

the measurement ranges of the sensor

e. Clean the sensor according to the guidelines described in the

Maintenance section

f. Ensure that the minimum immersion depth is maintained or incorrect

conductivity and/or temperature readings can result


Technical Support

a. Call us for technical support: 800-876-1172 or (979) 690-5560 (many

problems can be solved over the phone). Fax: (979) 690-0440 or

Email: [email protected].

b. In the event that the equipment needs to be returned to the factory for

any reason, please call to obtain a RMA # (Return Material