Operation
VARIO Cond
42
Resetting the
measuring param-
eters
Setting the
calibration interval
1
Start the INIT initialization menu with <
C_
>.
2
For initialization, select YES with <
§
>.
3
Confirm with <
g
>.
The reset is performed. The measuring instrument switches to the
setting of the calibration interval (see section 4.8).
The display flashes.
STO
RCL
C
S/cm
Tref25
nLF
MEM
CAL
TDS
CellTC
S
SAL
O
2
TC
RESET
Cell
LAP
m
mbar
mg/l
CAL
LoBat Day.Month h:min min:s
Ident
Lap
8
CLR
Lin
Tref20
Tref20
M cm
k cm
Sal
1
cm
K
ESC
1 n i t
Y E S
4
Terminate the setting with <
E
>.
or
Set the calibration interval in days (d) with <
<
> and <
>
>.
5
Confirm the setting with <
g
>.
6
Switch to the measuring mode with <
g
>.
Day.Month
MEM
CAL
d
C
pH
8
LoBat
min:s
Ident
Lap
mV
h:min
CLR
RESET
LAP
ESC
ESC
I n t
0
8
d 1