ProLab 3000
Ion concentration
115
ba75645e02
01/2010
3
In the Sample register, enter the values for the sample volume
(Sample volume) and the volume of the ISA/TISAB solution
(ISA/TISAB volume) as necessary.
4
Thoroughly rinse the electrode with deionized water.
5
Immerse the electrode in the sample.
6
Start the measurement with [Continue] .
After end of the measurement, the Standard pops up.
7
In the Standard register, enter the value for Standard concen-
tration if necessary.
8
Add standard solution to the sample (1 percent by volume of
the sample solution).
9
Start the measurement with [Continue].
The measured value is checked for stability (Stability control).