Xylem photoLab 6xxx Car Adapter User Manual

Page 4

Advertising
background image

Car Adapter photoLab

®

6xxx

4

ba75758de03

08/2009

Safety instructions

For operation with an external battery, follow the safety instructions of the
battery.
Make sure the power source is suitable for operation of the spectrophotome-
ter (see technical data of the power source and technical data of the spectro-
photometer).

Operating time

with a battery

The maximum operating time depends on various factors:

z Battery (e.g. nominal capacity, condition, age)

z Operating mode of the photometer (e.g. frequency of measurements)

z Photometer (instrument type)

Example:

Operating time

with a battery

of the type

12 V / 19 Ah

Note
The spectrophotometer consumes electricity even while it is in standby
mode.
We recommend to disconnect the Car Adapter photoLab

®

6xxx if you do not

use the spectrophotometer during battery operation.

Technical data

Car Adapter

photoLab

®

6xxx

Instrument type

Operating time with the battery
in optimum condition

photoLab

®

6100 UV

approx. 6.5 h

photoLab

®

6600 UV-VIS

approx. 16 h

Cable length

2 m

Max. voltage

12 V

Max. current

8 A

Advertising