MPP 350

What to do if...

23

ba75434e02

07/2005

6.2

Error symptoms of conductivity measurement

6.3

Error symptoms of pH measurement

Error symptom

Cause

Remedy

No temperature or con-
ductivity display

– No connection between meter and

sensor

– Cable defective

– Check connection between meter

and sensor

– Return the sensor

Measurement delivers
implausible conductivity
values

– Incorrect cell constant adjusted at

the meter

– Measuring range exceeded
– Contamination in the area of the

electrodes

– Electrodes damaged

– Check / correct the cell constant
– Observe the application range
– Clean the sensor (see

section 5.2).

– Return the sensor

Incorrect temperature
display

– The temperature sensor was not

immersed in the measuring
solution

– Temperature sensor defective

– Observe the minimum immersion

depth

– Return the sensor

Error symptom

Cause

Remedy

No pH display

– No connection between meter and

sensor

– Cable defective
– pH electrode defective

– Check connection between meter

and sensor

– Return the sensor
– Replace pH electrode