Xylem MPP 350 User Manual

Page 27

Advertising
background image

MPP 350

Technical data

27

ba75434e02

07/2005

Measurement

conditions

Measuring ranges at 20 °C

0 ... 50 mg/l D.O.
0 ... 600 % D.O. saturation
0 ... 1250 mbar D.O. partial pressure

Polarization time

min. 30 ... 50 min after changing the electro-
lyte,
it is not necessary to connect the sensor to
the meter for polarization

Temperature range

0 ... 50 °C

Max. allowed overpressure

10

6

Pa (10 bar)

Depth of immersion with
conductivity basic module

min. 130 mm (with protective hood and sink-
er)
max. 100 m (depending on the cable length)

Operating position

any

Approach flow

> 3 cm/s at 10 % measurement accuracy
10 cm/s at 5 % measurement accuracy
18 cm/s at 1 % measurement accuracy

Storage

conditions

Recommended storing
method

in the OxiCal

®

-CX calibration vessel (moist-

ened)

Storage temperature

0 ... 50 °C

Characteristic data

on delivery

Zero signal

< 0.1 % of the saturation value

Response time at 20 °C

t

90

(90 % of the final value display after)

< 10 s
t

95

(95 % of the final value display after)

< 16 s
t

99

(99 % of the final value display after)

< 60 s

Own consumption of oxy-
gen at 20 °C

0.008

µg·h

-1

(mg/l)

-1

Drift

approx. 3 % per month in the operating con-
dition

Working life

min. 6 months with one electrolyte filling