LR 925/01

Cleaning

13

ba75825e01

08/2009

2

Cleaning

CAUTION

To clean the sensor, disconnect it from the instrument.

Outside cleaning

We recommend to clean the sensor thoroughly, especially before measuring
low conductivity values.

After cleaning, thoroughly rinse with deionized water and recalibrate if
necessary.

Aging of the

conductivity

measuring cell

Normally, the conductivity measuring cell does not age. Special measuring
media (e.g. strong acids and bases, organic solvents) or temperatures that
are too high may considerably reduce its lifetime or lead to damage. The
warranty does not cover failure caused by measuring conditions and
mechanical damage.

Disposal

We recommend to dispose of the measuring cell as electronic waste.

3

What to do if...

Contamination

Cleaning procedure

Lime sediments

Immerse in acetic acid for 5 minutes (volume
share = 10 %)

Fat/oil

Clean with warm water that contains
washing-up liquid

Error symptom

Cause

Remedy

No temperature or
conductivity display

– No connection between meter and

conductivity measuring cell

– Cable defective

– Establish connection between

meter and conductivity measuring
cell

Measurement delivers
implausible conductivity
values

– Measuring range exceeded

– Contamination in the area of the

electrodes

– Electrodes damaged

– Make sure the correct sensor is

being used for the application

– Clean the conductivity measuring

cell (see section 2).

– Return the sensor

Incorrect temperature
display

– The temperature sensor is not

immersed deep enough in the
measuring solution

– Temperature sensor defective

– Observe the minimum immersion

depth

– Return the conductivity measuring

cell