4 technical data, 1 general data, Technical data – Xylem LR 925/01 User Manual
Page 6: General data, 4technical data

Technical data
LR 925/01
14
ba75825e01
08/2009
4
Technical data
4.1
General data
General features
Measuring principle
2-electrodes measurement
Cell constant
0.100 cm
-1
± 2 %
Temperature sensor
Integrated NTC 30 (30 k
:
at 25 °C / 77 °F)
Dimensions
(in mm)
Weight
Approx. 90 g (without cable)
Materials
Shaft
Stainless steel 1.4571
Connection head
POM
Conductivity electrodes
Stainless steel 1.4571
Thermistor enclosure
Stainless steel 1.4571
Connection cable
Lengths
1.5 m
Diameter
4.3 mm
Smallest allowed
bend radius
Fixed installation:
20 mm
Flexible use:
60 mm
Plug type
Socket, 4 pins
Pressure
resistance
Sensor with connection
cable
IP 68 (2 x 10
5
Pa or 2 bar)
Cable plug
IP 67 (when plugged in)
The LR 925/01 meets the requirements according to article 3(3) of the
directive, 97/23/EC ("pressure equipment directive").
120.0
12.0
176.0
21.7