Technical data, General data, Measurement and application characteristics – Xylem Sentix 980 User Manual

Page 4: Connection cable, Accuracy of the ids measuring technique

Advertising
background image

Technical data

SenTix

14

ba75843e01

08/2009

Technical data

General data

Measurement and application characteristics

Shaft dimensions, shaft material, electrical connection

Connection cable

Accuracy of the IDS measuring technique

WTW model

Reference electrolyte

Junction

NTC

Special features

SenTix

®

950

KCl 3 mol/L, Ag

+

free

Ceramic

Yes

Plastic shaft

SenTix

®

980

KCl 3 mol/L, Ag

+

free

Platinum

Yes

WTW model

pH measuring range

Allowed temperature
range

SenTix

®

950

0.000 ... 14.000

0 ... 80 °C (32 ... 176 °F)

SenTix

®

980

0.000 ... 14.000

0 ... 100 °C
(32 ... 212 °F)

WTW model

Membrane resistance
at 25 °C

Typical application

SenTix

®

950

< 1 GOhm

Field

SenTix

®

980

< 600 MOhm

Laboratory

WTW model

Shaft

Electrical connection

Length
[mm]

Ø

[mm]

Material

Electrode
connection

Meter con-
nection

Cable
length

SenTix

®

950

120

12

Polyamide

Fixed cable

Digital plug

1.5 m

SenTix

®

980

120

12

Glass

Fixed cable

Digital plug

1.5 m

Diameter

4.3 mm

Smallest allowed bend radius

Fixed installation:

20 mm

Flexible use:

60 mm

Plug type

Socket, 4 pins

Measured parameter

Accuracy (± 1 digit)

pH

± 0.004

U [mV]

± 0.2

T [°C]

± 0.1