Cleaning

Remove water-soluble contamination by rinsing with deionized water. Re-
move other contamination as follows:

Note
Hydrofluoric acid, hot phosphoric acid and strong alkaline solutions destroy
the glass membrane.

After cleaning

Rinse the electrode with deionized water and condition it in reference
electrolyte solution for at least 1 hour. Then recalibrate the electrode.

Contamination

Cleaning procedure

Fat and oil

Rinse with water containing household wash-
ing-up liquid

Lime and hydroxide deposits Rinse with citric acid (10 % by weight)

Proteins

Immerse in pepsin cleaning solution PEP/pH
for approx. 1 hour.
Note: Make sure the level of the reference
electrolyte is above that of the cleaning solu-
tion.