Xylem Multi 3420 User Manual
Page 62
 
Operation
Multi 3420
62
ba75864e11
02/2011
The QSC initial calibration is completed. The sensor is calibrated. If you 
want to calibrate with special buffers for your measurements, you can 
subsequently carry out a normal calibration with your buffers. The ref-
erence values determined with the QSC calibration are also used for 
the evaluation of normal calibrations. In the measured value display, 
the color scale of the QSC function is always displayed. A double arrow 
on the color scale indicates the current sensor evaluation.
You can carry out QSC control calibrations at greater intervals than 
normal calibrations. 
A QSC control calibration can, e.g. be useful if the sensor evaluation 
noticeably changed (after some normal calibrations).
Carrying out a QSC
control calibration
4
As soon as the three-point calibration has been successfully 
carried out you can decide whether to accept or discard the cal-
ibration as the QSC initial calibration.
22.09.2009 08:00
Info
QSC color scale
The double arrow on 
the color scale indi-
cates the current sen-
sor evaluation.
1
Open the menu for measurement settings with <MENU/
ENTER>.
2
In the QSC menu, select Control calibration with <
S><T>.
The calibration display appears. AutoCal QSC-Kit is displayed 
as the buffer.
Exclusively use the QSC-Kit for the QSC calibration. If you use 
other buffers, you will have no valid QSC control calibration.
3
Follow the user guide.
The calibration is carried out like a normal three-point calibra-
tion. As soon as the three-point calibration has been success-
fully carried out you can decide whether to accept or discard the 
calibration as the QSC control calibration.