Operation
Multi 3420
80
ba75864e11
02/2011
Note
You can prematurely terminate the Stability control function manually 
with <MENU/ENTER> at any time. If the Stability control function is pre-
maturely terminated, the current measurement data are output to the 
interface without the AutoRead info.
Criteria for a stable
measured value
The Stability control function checks whether the measured values are 
stable within the monitored time interval. 
The minimum duration until a measured value is assessed as stable is 
the monitored time interval. The actual duration is mostly longer.
1
Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
2
Using <MENU/ENTER>, activate the Stability control function 
manually.
The [AR] status indicator appears while the measured value is 
assessed as not stable. A progress bar is displayed and the 
display of the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable 
measured value is recognized. The progress bar disappears 
and the display of the measured parameter stops flashing. 
The current measurement data is output to the interface. Mea-
surement data meeting the stability control criterion is marked 
by AR.
3
Using <MENU/ENTER>, start a further measurement with Sta-
bility control.
or
Release the frozen measured value again with <AR>.
The display switches to the measured value display. 
The [AR][HOLD] status display disappears. 
Measured 
parameter
Time interval
Stability in the time interval
Conductivity
ϰ
10 seconds
Δ
ϰ: better than
0.5% of measured value
Temperature
10 seconds
Δ
T (° C): Better than 0.02