Description
Specification
Supplemental information
Crosstalk
1
- standard
45 MHz to 1 GHz
85 dB
1 to 2 GHz
100 dB
2 to 20 GHz
110 dB
20 to 40 GHz
108 dB
40 to 45 GHz
105 dB
45 to 50 GHz
100 dB
Crosstalk
1
- Option UNL or 014
45 MHz to 1 GHz
85 dB
1 to 2 GHz
100 dB
2 to 20 GHz
109 dB
20 to 40 GHz
106 dB
40 to 45 GHz
103 dB
45 to 50 GHz
98 dB
Crosstalk
1
- Option UNL and 014
45 MHz to 1 GHz
85 dB
1 to 2 GHz
98 dB
2 to 10 GHz
108 dB
10 to 20 GHz
107 dB
20 to 40 GHz
104 dB
40 to 45 GHz
100 dB
45 to 50 GHz
95 dB
1. Measurement conditions: Normalized to a thru,
measured with two shorts, 10-Hz IF bandwidth,
averaging factor of 8, alternate mode, source power
set to the lesser of the maximum power out or the
maximum receiver power.
Advertising