Test port input – Atec Agilent-E8364A User Manual
Page 18

Test port input
continued
Description
Specification
Supplemental information
Standard
014
UNL
UNL and 014
Reference level magnitude
Range
±200 dB
±200 dB
±200 dB
±200 dB
Resolution
0.001 dB
0.001 dB
0.001 dB
0.001 dB
Reference level phase
Range
±500°
±500°
±500°
±500°
Resolution
0.01°
0.01°
0.01°
0.01°
Stability magnitude
1
Typical ratio measurement:
Measured at the test port
45 MHz to 20 GHz
±0.02 dB/°C
20 to 40 GHz
±0.03 dB/°C
40 to 50 GHz
±0.04 dB/°C
Stability phase
1
Typical ratio measurement:
Measured at the test port
45 MHz to 20 GHz
±0.2°/°C
20 to 40 GHz
±0.5°/°C
40 to 50 GHz
±0.8°/°C
Damage input level
Test port 1 and 2
20 dBm or ±40 VDC, typical
R1, R2 in
15 dBm or ±15 VDC, typical
A, B in
15 dBm or ±15 VDC, typical
Coupler thru (option 014 or UNL and 014)
30 dBm or ±40 VDC, typical
Coupler arm (option 014 or UNL and 014)
30 dBm or ±7 VDC, typical
18
1. Stability is defined as a ratio measurement
measured at the test port.