Test port input, Dynamic accuracy, Continued – Atec Agilent-E8364A User Manual
Page 20

Test port input
continued
Dynamic accuracy
Applies to input ports 1 and 2, accuracy of the test
port input power reading relative to the reference
input power level. Also applies to the following
conditions:
• IF bandwidth = 10 Hz
• test port powers = > –50 dBm and < 0 dBm
• magnitude dynamic accuracy = 0.02 dB +
0.001 dB/dB from reference power
• phase dynamic accuracy = 0.132° =
0.0066°/dB from reference power
20
Magnitude
0.01
0.1
1
10
-120
-110
-100
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
Testport power (dBm)
Accuracy (dB)
E8362/3/4A
■
-10 dBm at 1 GHz
◆
-20 dBm at 1 GHz
▲
-30 dBm at 1 GHz
●
-40 dBm at 1 GHz
Phase
0.1
1
10
100
-120
-110
-100
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
Testport power (dBm)
Accuracy (degrees)
E8362/3/4A
■
-10 dBm at 1 GHz
◆
-20 dBm at 1 GHz
▲
-30 dBm at 1 GHz
●
-40 dBm at 1 GHz