To remove or replace a system component – Atec Agilent-85133F User Manual

Page 103

Advertising
background image

Servicing the System

4

Installation and User’s Guide

105

To remove or replace a system component

Follow these steps to remove a system component for periodic component
calibration, service, or repair.

Follow the steps in reverse order to replace a system component.

WA R N I N G

These servicing instructions are provided for use by qualified
personnel only. To avoid electrical shock, do not perform any
servicing unless you are qualified to do so.

WA R N I N G

The opening of covers or removal of parts is likely to expose
dangerous voltages. Disconnect the product from all voltage
sources before opening covers or removing parts.

To remove a system component

Step

Note

1 Turn off all components.

2 Disconnect the system from line

power.

3 Disconnect all cables from the

instrument to be removed.

4 Prepare a surface near the rack

cabinet to place the removed
system component.

5 Lower the stabilizing legs on the

rack cabinet.

6 From the front of the rack cabinet,

remove the four screws (2 screws
on each rack mount) attaching the
instrument’s rack mount and
handle kit to the rack cabinet.

7 As another person steadies the

rack cabinet, hold the instrument
by the rack mount handles and
slowly pull forward.

•

Some components may weigh more
than 50 pounds and may require
more than one person to remove
safely.

8 Save the rack mount screws by

reinserting them in the rack cabinet
frame nuts.