Installation and User’s Guide

3

Verifying System Functionality

3 Start the IC-CAP

software.

*

PC version:

a Select Start > Programs > IC-CAP 2004 >

IC-CAP.

UNIX version:

a Open a UNIX terminal window.
b At the prompt, type

iccap

.

c Press Enter.

•

This starts IC-CAP and opens the
IC-CAP/Status and IC-CAP/Main
windows.

4 Add the system interface

and components to the
IC-CAP Hardware Setup.

a From the IC-CAP/Main window menu bar,

choose Tools > Hardware Setup... .

b If you have not added an interface, click

Add Interface, enter the system interface
(

hpib

or other), and click OK.

c In the Instrument List, click Rebuild.
d Verify that all system components appear

in the Instrument List.

•

This opens the IC-CAP/Hardware
Setup window.

•

This polls the GPIB and adds all
connected and activated system
instruments to the Instrument List.

•

Disregard error messages on the
system instrument displays. The errors
are a by-product of the GPIB polling
process.

5

For 4156C systems,
change the Agilent 4156C
precision semiconductor
parameter analyzer
HRSMU names in IC-CAP.

a In the Instrument List, select HP4156 (hpib,

19) and click Configure... .

b In the Configuration of HP4156 window

Unit Table group box, highlight the
characters in the HRSMU1 entry box and
type

VG

.

c In the Unit Table group box, highlight the

characters in the HPSMU entry box and
type

VD

.

d Click OK.

†

e Close the Hardware Setup window.

•

This opens the Configuration of
HP4156 window where the Unit Table
dialog box is used to change the names
of the HRSMUs.

•

Actions b, c, and d change the name of
HRSMU1 to VG and HPSMU to VD.

•

Renaming the SMUs is necessary for
proper execution of the example model
file.

6

For E5260A/70B systems,
change the Agilent 4156C
precision semiconductor
parameter analyzer SMU
names in IC-CAP.

a In the Instrument List, select Agilent E5270

(hpib, 19) and click Configure... .

b In the Configuration of E5270 window Unit

Table group box, highlight the characters in
the MPSMU<slot number> entry box and
type

VG

.

c In the Unit Table group box, highlight the

characters in the HPSMU<slot number>
entry box and type

VD

.

d Click OK.

‡

e Close the Hardware Setup window.

•

This opens the Configuration of E5270
window where the Unit Table dialog
box is used to change the names of the
SMUs.

•

Actions b, c, and d change the name of
MPSMU<slot number> to VG and
HPSMU<slot number> to VD.

•

Renaming the SMUs is necessary for
proper execution of the example model
file.

To perform the system functional verification test (continued)

Step

Action

Notes

Advertising