Power range, Vectorstar™ ms4640b series power range – Atec Anritsu-MS4640B User Manual
Page 7

VectorStar™ MS4640B Series
Power Range
MS4640B Series TDS
PN: 11410-00611 Rev. A
7
Power Range
Maximum Rated Power to minimum level. The difference reflects the ALC range for standard models or with Option 051, and the ALC + Attenuator Range
for models with Options 061 or 062. Maximum Rated Power is typical from 2.4 - 2.7 GHz.
MS4642B, 20 GHz Model, Power Range (dBm)
MS4644B, 40 GHz Model, Power Range (dBm)
MS4645B & MS4647B, 50 & 70 GHz Models, Power Range (dBm)
Frequency
Standard
Option 051
Option 061
a
or 062
a. The Option 061 power range reported in this column applies to Port 1. For Port 2, use the figures from the Option 051 column.
70 kHz to 0.01 GHz
+10 to –25
+9 to –25
+8 to –95
0.01 to 2.5 GHz
+12 to –25
+11 to –25
+10 to –95
2.5 to 20 GHz
+13 to –20
+12 to –20
+11 to –90
With Option 031
70 kHz to 0.01 GHz
+12 to –25
+11 to –25
+10 to –95
0.01 to 2.5 GHz
+14 to –25
+13 to –25
+12 to –95
2.5 to 20 GHz
+14 to –20
+13 to –20
+12 to –90
Frequency
Standard
Option 051
Option 061
a
or 062
a. The Option 061 power range reported in this column applies to Port 1. For Port 2, use the figures from the Option 051 column.
70 kHz to 0.01 GHz
+10 to –25
+9 to –25
+8 to –95
0.01 to 2.5 GHz
+12 to –25
+11 to –25
+10 to –95
2.5 to 20 GHz
+9 to –20
+8 to –20
+7 to –90
20 to 40 GHz
+9 to –25
+8 to –25
+7 to –95
With Option 031
70 kHz to 0.01 GHz
+12 to –25
+11 to –25
+10 to –95
0.01 to 2.5 GHz
+14 to –25
+13 to –25
+12 to –95
2.5 to 20 GHz
+12 to –20
+11 to –20
+10 to –90
20 to 40 GHz
+12 to –25
+11 to –25
+10 to –95
Frequency
Standard
Option 051
Option 061
a
or 062
a. The Option 061 power range reported in this column applies to Port 1. For Port 2, use the figures from the Option 051 column.
70 kHz to 0.01 GHz
+10 to –25
+9 to –25
+8 to –85
0.01 to 2.5 GHz
+12 to –25
+11 to –25
+10 to –85
2.5 to 5 GHz
+6 to –20
+5 to –20
+3 to –80
5 to 20 GHz
+5 to –20
+4 to –20
+2 to –80
20 to 38 GHz
+6 to –25
+4 to –25
+2 to –85
38 to 50 GHz
+5 to –25
+3 to –25
+1 to –85
50 to 65 GHz
0 to –25
–2 to –25
–4 to –85
65 to 67 GHz
–2 to –25
–3 to –25
–5 to –85
67 to 70 GHz
–3 to –25
–6 to –25
–10 to –85
With Option 031
b
b. With Option 084 or 085, Test Port 2 maximum power is equivalent to the non-option 031 range (typical). 38 to 50 GHz range may degrade
by up to 3 dB.
70 kHz to 0.01 GHz
+12 to –25
+11 to –25
+10 to –85
0.01 to 2.5 GHz
+14 to –25
+13 to –25
+12 to –85
2.5 to 5 GHz
+8 to –20
+7 to –20
+5 to –80
5 to 20 GHz
+8 to –20
+7 to –20
+5 to –80
20 to 38 GHz
+8 to –25
+6 to –25
+4 to –85
38 to 50 GHz
+7 to –25
+5 to –25
+3 to –85
50 to 65 GHz
7 to –25
+5 to –25
+3 to –85
65 to 67 GHz
+6 to –25
+4 to –25
+2 to –85
67 to 70 GHz
+4 to –25
+1 to –25
–3 to –85