Power range, Vectorstar™ ms4640b series power range – Atec Anritsu-MS4640B User Manual

Page 7

Advertising
background image

VectorStar™ MS4640B Series

Power Range

MS4640B Series TDS

PN: 11410-00611 Rev. A

7

Power Range

Maximum Rated Power to minimum level. The difference reflects the ALC range for standard models or with Option 051, and the ALC + Attenuator Range

for models with Options 061 or 062. Maximum Rated Power is typical from 2.4 - 2.7 GHz.

MS4642B, 20 GHz Model, Power Range (dBm)

MS4644B, 40 GHz Model, Power Range (dBm)

MS4645B & MS4647B, 50 & 70 GHz Models, Power Range (dBm)

Frequency

Standard

Option 051

Option 061

a

or 062

a. The Option 061 power range reported in this column applies to Port 1. For Port 2, use the figures from the Option 051 column.

70 kHz to 0.01 GHz

+10 to –25

+9 to –25

+8 to –95

0.01 to 2.5 GHz

+12 to –25

+11 to –25

+10 to –95

2.5 to 20 GHz

+13 to –20

+12 to –20

+11 to –90

With Option 031

70 kHz to 0.01 GHz

+12 to –25

+11 to –25

+10 to –95

0.01 to 2.5 GHz

+14 to –25

+13 to –25

+12 to –95

2.5 to 20 GHz

+14 to –20

+13 to –20

+12 to –90

Frequency

Standard

Option 051

Option 061

a

or 062

a. The Option 061 power range reported in this column applies to Port 1. For Port 2, use the figures from the Option 051 column.

70 kHz to 0.01 GHz

+10 to –25

+9 to –25

+8 to –95

0.01 to 2.5 GHz

+12 to –25

+11 to –25

+10 to –95

2.5 to 20 GHz

+9 to –20

+8 to –20

+7 to –90

20 to 40 GHz

+9 to –25

+8 to –25

+7 to –95

With Option 031

70 kHz to 0.01 GHz

+12 to –25

+11 to –25

+10 to –95

0.01 to 2.5 GHz

+14 to –25

+13 to –25

+12 to –95

2.5 to 20 GHz

+12 to –20

+11 to –20

+10 to –90

20 to 40 GHz

+12 to –25

+11 to –25

+10 to –95

Frequency

Standard

Option 051

Option 061

a

or 062

a. The Option 061 power range reported in this column applies to Port 1. For Port 2, use the figures from the Option 051 column.

70 kHz to 0.01 GHz

+10 to –25

+9 to –25

+8 to –85

0.01 to 2.5 GHz

+12 to –25

+11 to –25

+10 to –85

2.5 to 5 GHz

+6 to –20

+5 to –20

+3 to –80

5 to 20 GHz

+5 to –20

+4 to –20

+2 to –80

20 to 38 GHz

+6 to –25

+4 to –25

+2 to –85

38 to 50 GHz

+5 to –25

+3 to –25

+1 to –85

50 to 65 GHz

0 to –25

–2 to –25

–4 to –85

65 to 67 GHz

–2 to –25

–3 to –25

–5 to –85

67 to 70 GHz

–3 to –25

–6 to –25

–10 to –85

With Option 031

b

b. With Option 084 or 085, Test Port 2 maximum power is equivalent to the non-option 031 range (typical). 38 to 50 GHz range may degrade

by up to 3 dB.

70 kHz to 0.01 GHz

+12 to –25

+11 to –25

+10 to –85

0.01 to 2.5 GHz

+14 to –25

+13 to –25

+12 to –85

2.5 to 5 GHz

+8 to –20

+7 to –20

+5 to –80

5 to 20 GHz

+8 to –20

+7 to –20

+5 to –80

20 to 38 GHz

+8 to –25

+6 to –25

+4 to –85

38 to 50 GHz

+7 to –25

+5 to –25

+3 to –85

50 to 65 GHz

7 to –25

+5 to –25

+3 to –85

65 to 67 GHz

+6 to –25

+4 to –25

+2 to –85

67 to 70 GHz

+4 to –25

+1 to –25

–3 to –85

Advertising