Specifi cations – Atec Agilent-6570A Series User Manual

Page 2

Advertising
background image

2

Specifi cations

Specifications
(at 0 ° to 55 °C unless otherwise specified)

6571A

6572A

6573A

6574A

6575A

6571A-J03

Special order

option

Number of outputs

1 1 1 1 1

1

GPIB

No No No No No

No

Output ratings

Output voltage

0 to 8 V

0 to 20 V

0 to 35 V

0 to 60 V

0 to 120 V

14 V

Output current

0 to 220 A

0 to 100 A

0 to 60 A

0 to 35 A

0 to 18 A

150 A

Programming accuracy (at 25 °C ± 5 °C)

Voltage

0.04% +

8 mV

20 mV

35 mV

60 mV

120 mV

14 mV

Current

0.1% +

125 mA

60 mA

40 mA

25 mA

12 mA

90 mA

Ripple and noise from 20 Hz to 20 MHz

Voltage

rms

650 μV

750 μV

800 μV

1.25 mV

1.9 mV

1.5 mV

peak-to-peak

7 mV

9 mV

9 mV

11 mV

16 mV

15 mV

Current

rms

200 mA

100 mA

40 mA

25 mA

12 mA

150 mA

Load and line regulation

Voltage

0.002% +

300 μV

650 μV

1.2 mV

2 mV

4 mV

600 μV

Current

0.005% +

10 mA

7 mA

4 mA

2 mA

1 mA

7 mA

Transient response time

Less than 900 μs for the output voltage to recover to within 100 mV following a change
in load from 100% to 50% or 50% to 100% of the output current rating of the supply

Supplemental characteristics
(Non-warranted characteristics determined
by design and useful in applying the product)

6571A

6572A

6573A

6574A

6575A

6571A-J03

Special order

option

Average programming resolution

Voltage

2 mV

5 mV

9 mV

15 mV

30 mV

4 mV

Current

55 mA

25 mA

15 mA

8.75 mA

4.5 mA

40 mV

OVP

15 mV

35 mV

65 mV

100 mV

215 mV

28 mV

Output voltage programming
response time*
(excluding command
processing time)

30 ms

60 ms

130 ms

130 ms

195 ms

30 ms

* Full load programming rise/fall time (10% to 90% or 90% to 10%) with full resistive load equal to rated output voltage/rated output current.

Advertising