Smartwave – Atec Elgar-SW1750A-SW3500A-SW5250A User Manual

Page 4

Advertising
background image

4

4

OUTPUT

Output Power: 1750 VA: 1ø, 3500 VA: 1ø or 2ø,
5250 VA: 1ø or 3ø (systems up to 21,000 VA)

AC or DC Output Voltage: 0 to156 VRMS L-N, range
1; 0 to 312 VRMS L-N, range 2

Voltage Accuracy: ±0.1% of range. Above 1 kHz,
add 0.2%/kHz. Add ±0.1% of full scale for “AC PLUS
DC” mode. Valid for 5 to 156 VRMS and 10 to 312
VRMS at 25°C

Voltage Resolution: 0.05% of full scale

Output Current Per Phase: 13A to 135V in 156V
range; 6.5A to 270V in 312V range; optional 16A/8A
consult factory (per 1750 VA module)

Note: Higher currents may be achieved in 10.5-21 kVA systems.

Crest Factor: 4.0 (peak output current to RMS
output current)

Power Factor of Load: 0 lagging to 0 leading

Output Frequency Range: DC or 40 Hz to 5 kHz. For
output frequencies greater than 1 kHz, the max slew
rate allowed is 1 kHz per second.

Frequency Accuracy: ±0.01% at 25°C
±0.001%/°C

Frequency Resolution:

40 Hz to 99.99 Hz: 0.01 Hz

100 Hz to 999.9 Hz: 0.05 Hz

1000 Hz to 5000 Hz: 0.5 Hz

Load Regulation: ±0.025% of full scale voltage
for a full resistive load to no load; above 1 kHz, add
±0.01%/kHz

Line Regulation (DC, or 40 Hz to 5 kHz): ±0.025%
of full scale for a ±10% input line change

Max Total Harmonic Distortion
(Full Linear Load or No Load): 0.25% max, 40 to
100 Hz; 0.5% max to 500 Hz; and 1% max to 1 kHz
plus 1%/kHz to 5 kHz

Phase Accuracy, Phase-to-Phase Balanced Linear
Resistive Load: ±1°, 40 Hz to 1 kHz, plus ±1°/kHz
above 1 kHz

Phase Angle Resolution: 0.1°

AC Noise Level: >60 dB RMS below full output
voltage

Amplitude Stability With Remote Sense: ±0.1%
of full scale over 24 hours at constant line, load and
temperature

Remote Output Voltage Sense: 5 VRMS total lead
drop, max

WAVEFORM SPECIFICATIONS

Waveshape Libraries: 50 factory supplied in ROM;
storage available for up to 50 user created in non-
volatile RAM

Front Panel Setups: A total of 50 user created
steady-state waveforms, with amplitude, frequency,
phase angle and current limit parameters

Sequencing Library: 1000 user created segments
stored in non-volatile RAM. Segments include
waveshape, amplitude, frequency, phase angle, time
(from 1 ms to 1000 seconds), or number of cycles.

MIL-STD-704 Transient Library

MEASUREMENTS (OPTIONAL)

• Phase to Neutral RMS Output Voltages

• Phase to Phase Voltages

• 1ø to 3ø RMS Output Currents

• Peak Current

• Output Frequency

• 1ø to 3ø Power

• 1ø to 3ø VA

• Power Factor of 1ø or 3ø Loads

• Output Phase Angles Relative to Phase A

Measurement Capability: 4.5 Digit Analog to
Digital Measurement System with .01% of full scale
resolution unless otherwise noted.

Calibration Interval: 12 months

Temperature Range for specifi ed Accuracy:
25°C to ±5°C unless otherwise noted

Phase to neutral RMS voltage measurement:
Valid for phases A, B and C (use phase A for Parallel
Mode)

• Range: 0V to 350V plus sign bit for DC range
• Accuracy: ±0.3% of range, DC or 47 Hz to 1
kHz; ±0.5% of range, 40 to 47 Hz and for 1 kHz
to 5 kHz

Phase to phase RMS voltage: Calculated from
Phase to Neutral voltages and phase angle

• Range: 0V to 700V
• Accuracy and Temperature Coeffi cient: The
same as the Phase to Neutral voltage

RMS current measurement: Valid for phases A, B,
and C (use phase A for Parallel Mode)

• Range 1: 0A to 7.5A, plus sign bit for DC

range; 3ø mode, 312V range

• Range 2: 0A to 15A, plus sign bit for DC

range; 3ø mode, 156V range

• Range 3: 0A to 22.5A, plus sign bit for DC

range; parallel mode, 312V range

• Range 4: 0A to 45A plus sign bit for DC

range; parallel mode, 156V range

• Accuracy: ±1.0% of range, DC or 40 Hz to

500 Hz; add ±1.5%/kHz above 500 Hz
Accuracies are specifi ed for a max crest factor
of 4.0

Note: Contact factory for measurement range on 10.5 to 21 kVA
system.

SMARTWAVE

TM

(SW)

SPECIFICATIONS

Advertising