Sweep range

Frequency sweep

Within instrument frequency range

Amplitude sweep

Within attenuator hold range

Dwell time

1 ms to 60 s

Frequency settling time

28 ms typical

Amplitude settling time

10 ms typical

Number of points

2 to 1601

Triggering

Auto, external, single, or GPIB

Output

Power

4

(dBm)

Frequency range

Standard

Option 1EA

20 GHz Models
250 kHz to 3.2 GHz

–20 to +13

–20 to +16

> 3.2 to 20 GHz

–20 to +13

–20 to +20

40 GHz Models
250 kHz to 3.2 GHz

–20 to +9

–20 to +15

> 3.2 to 20 GHz

–20 to +9

–20 to +18

> 20 to 40 GHz

–20 to +9

–20 to +14

20 GHz Models with option 1E1
250 kHz to 3.2 GHz

–135 to +11

–135 to +15

> 3.2 to 20 GHz

–135 to +11

–135 to +18

40GHz Models with option 1E1
250 kHz to 3.2 GHz

–135 to +7

–135 to +14

> 3.2 to 20 GHz

–135 to +7

–135 to +16

> 20 to 40 GHz

–135 to +7

–135 to +12

Option 1E1 step attenuator

0 dB and 5 to 115 dB in 10 dB steps

Attenuator hold range
(Same as max power sweep range)

Minimum

From –20 dBm to maximum specified output power.
Can be offset using Option 1E1 attenuator.

Amplitude switching speed

5

CW or analog modulation

< 25 ms, typical

When using power search

< 25 ms, typical

CW level accuracy

6

(dB)

Frequency

> +10 dBm

+10 to –10 dBm

–10 to –20 dBm

250 kHz to 2 GHz

±0.6

±0.6

±1.4

2 GHz to 20 GHz

±0.8

±0.8

±1.2

> 20 to 40 GHz

±1.0

±0.9

±1.3

4

Maximum power specification is warranted from
15 to 35ºC, and is typical from 0 to 15ºC. Maximum
power over the 35 to 55ºC range typically degrades
less than 2 dB.

5

To within 0.1 dB of final amplitude within one
attenuator range

6

Specifications apply over the 15 to 35ºC temperature
range. Degradation outside this range, for power levels
> -10 dBm, is typically < 0.3 dB.

For instruments with type-N connectors (Option 1ED),
specifications are degraded typically 0.2 dB above
18 GHz.