6 Fluke Corporation Fluke 430 Series Three-Phase Power Quality Analyzers

Technical specifications

The specifications of the instrument are verified using the “implementation verification”

table 3 as specified in IEC 61000-4-30 2002 Chapter 6.2. Accuracy is specified in % of

reading unless otherwise specified. Specifications are valid for models Fluke 435 and

Fluke 434 unless otherwise specified.

Input characteristics

Voltage inputs
Number of inputs

4 (3 phases + neutral) dc-coupled

Maximum input voltage

1000 Vrms

Nominal voltage range

50 V to 500 V according IEC 61000-4-30. Selectable from 1 V to 1000 V.

Maximum peak measurement voltage

6 kV

Input impedance

4 M

Ω//5 pF

Bandwidth

> 10 kHz, up to 100 kHz for transient display

Scaling

1:1, 10:1, 100:1, 1000:1 and variable

Current inputs
Number of inputs

4 (3 phases + neutral) dc-coupled

Type

Clamp on current transformer with mV output

Range

1 Arms to 400 Arms with included clamps (i400s/Fluke 434)

30 A to 3000 Arms with included clamps (i430-flex/Fluke 435)

1 Arms to 3000 Arms with optional clamps

Input impedance

50 k

Ω

Bandwidth

> 10 kHz

Scaling

0.1, 1, 10, 100, 1000 mV/A, variable, i5s and i430-flex

Nominal frequency

40 Hz to 70 Hz

Sampling system
Resolution

16 bit analog to digital converter on 8 channels

Maximum sampling speed

200 kS/s on each channel simultaneously

RMS sampling

5000 samples on 10/12

2

cycles according IEC 61000-4-30

PLL synchronization

4096 samples on 10/12

2

cycles according IEC 61000-4-7

Display modes

Waveform display

Available in Scope and Transient mode

Captures 8 waveforms simultaneously

Display update rate 5x per second

Up to 10/12

2

times horizontal zoom

Cursors: single vertical line showing min, max, avg reading at cursor position

Phasor

Shows real time phasor diagram

Available in Scope and Unbalance mode

Display update rate 5x per second

Meter readings

Available in Volts/Amps/Hertz, Harmonics, Power and Energy, Flicker, Unbalance and Logger

4

mode

AutoTrend graph

Available in Volts/Amps/Hertz, Dips and Swells, Harmonics, Power and Energy, Flicker, Unbalance, Inrush,

Mains Signaling

4

Logger

4

and Monitor mode

Cursors: single vertical line showing with min, max, avg reading at cursor position

Bargraph

Available in Harmonics and Monitor mode

Event list

Available in Dips and Swells, Mains Signaling

4

, Logger

4

and Monitor mode