SPECIFICATIONS

OUTPUT VOLTAGE:

0 to 3 5 VDC, coarse and fine

adjustment.

OUTPUT CURRENT:

0

to

10

A,

coarse

and

fine

adjustment.

CONSTANT VOLTAGE OPERATION

Voltage Regulation
Line (108 -132 V):

≤0.2% +2 mV

Load (no load to full load):

≤0.04% + ≤0.04% + 2 mV

Recovery Time:

≤100 μs typical.

Ripple Voltage
RMS:

≤1 mV.

Temperature Coefficient

(0° to +40 °C):

≤300 PPM/°C

CONSTANT CURRENT OPERATION

Adjustable Current Limits:

5% to 100%

Current Regulation
Line (108 -132 V):

≤0.4% + 5 mA

Load

≤0.4% + 5 mA

Current Ripple:

≤3 mA typical.

METERING
Model 1745A
Voltmeter

4 digit Green LED Display

Range:

0 to 99.99 V.

Accuracy:

(0.5% + 9 digits)

Ammeter

4 digit Red LED Display

Hi Range:

0 to 9.999 A.

Accuracy:

(0.5% + 9 digits).

Model 1744A
Voltmeter Range:

0 to 40 V

Accuracy:

+2.5%

Ammeter Hi Range:

0 to 11 A

Lo Range:

0 to 2.2 A.

Accuracy:

+2.5%

POWER REQUIREMENTS:
Domestic: 120 VAC ±10%, 60 Hz
International: 120/220/230/240 VAC - 10%, 50/60 Hz

POWER CONSUMPTION:

Approximately 420 W or less at
full load.

PROTECTION:

Reverse polarity protection,
Current limiting.

6