( H ) NO CURRENT CONTROL.

NO CURRENT CONTROL

Is Current Pot Open ?

Replace it.

Are Wires related to Current Pot OK ?

Ensure proper Connectivity.

1.

Check the Current Potentiometer. It might be Open.

2.

Check all Wires related to Current Potentiometer.

Y

N

Y

( I )

POOR LINE REGULATION.

POOR LINE REGULATION

Improper Device Drop Setting.

Keep the Input Supply at rated I/P.
Keep the O/P Supply at rated Vtg.
and rated Load. Measure Vtg.
across O/P +Ve and Capacitor

Leaky Summing Capacitor C4.

+Ve Adjust that Vtg. between 4.5V
to 5V with Preset VR1 ( 5K ) on
PREG PCB.

Replace It.

1.

Check whether Device Drop is between 4.5V to 5V. It might not be proper.

2.

Leaky Summing Capacitor C4 on PREG PCB.

Y

N

Y

23

Advertising