Controls and indicators – B&K Precision 1655A - Manual User Manual

Page 8

Advertising
background image

6

FRONT PANEL CONTROLS

(Refer to Fig. 1 and 2)

1.

POWER ON Switch.

2.

POWER ON Pilot Light.

3.

Function Switch.

Model 1655A

Five interlocking pushbutton switches which select
function and meter scale. Pressing a button releases
the previous selection. The following selections are
available:

VOLTS. Connects meter to measure voltage at

ISOLATED OUTPUT. Use 0-150 VOLTS
scales on meter. Voltage is adjustable with AC
VOLTS control.

AMPS 0 – 4. Connects meter to measure current of

ISOLATED OUTPUT. USE 0-4 AMPS scale on
meter. When voltage is set to 120 VAC, the 0-480
VA scale may be used to measure the output load
in volt-amps



AMPS 0-2.
Connects meter to measure current of

ISOLATED OUTPUT. Use 0-2 AMPS scale on
meter. When voltage is set to 120VAC, the
0-240VA scale may be used to measure the
output load in volt-amps.


LEAKAGE COM.
Connects meter to measure leakage with

respect to common side of ac line. Use LEAKAGE scale on
meter. Voltage must be preset to 120VAC for accurate
leakage measurements.

LEAKAGE HOT. Connects meter to measure leakage with

respect to hot side of ac line. Use LEAKAGE scale on
meter. Voltage must be preset to 120VAC for accurate
leakage measurements.

Model 1653A

Alternate action pushbutton switch selects function of meter.
Push once to latch button in AMPS (in) position. Push again to
release button to VOLTS (out) position. The following
selections are available:
VOLTS. Connects meter to measure voltage at ISOLATED

OUTPUT. Use 0-150 VOLTS scale on meter. Voltage is
adjustable with AC Volts control.

AMPS. Connects meter to measure current of ISOLATED

OUTPUT. Use 0-2 AMPS scale on meter.

4.

Meter. Appropriate scale is selected by function switch.

5.

AC VOLTS Control. Adjusts voltage at ISOLATED

OUTPUT receptacle(s) from 0 to 150 volts.



CONTROLS AND INDICATORS