Elationship, Etween, Eference – B&K Precision 2658A - Manual User Manual

Page 33: Evel and, Att/amp, At d

Advertising
background image

27

* When the reference level is set below “Minimum”, “*S/W AMP” is displayed in Amplitude axis setting

values display area on the screen.

Calculation formula (conversion from dBm)

● A[dBμV] = 107 + X [dBm] ● B[dBmV] = 47 + X [dBm] ● C[dBV] = -13 + X [dBm]
● D[dBμV/m] = 68.8 / λ×√(X/Gar) [dBm] λ : Wavelength[m] Gar : Antenna absolute gain[times]
● E[dBμA/m] = 107 + X + F[dBm] F : Probe calibration coefficient[dB] **changes depending on the
frequency.

9.8 Relationship Between Reference Level and ATT/AMP (at dBm)

Internal input attenuator (ATT) and IF amplifier (AMP) are automatically set to the optimum values based on the

reference level (REFER). (The input attenuator cannot be set independently.)

REFER

(dBm)

ATT

(dB)

AMP

(dB)

REFER

(dBm)

ATT

(dB)

AMP

(dB)

REFER

(dBm)

ATT

(dB)

AMP

(dB)

REFER

(dBm)

ATT

(dB)

AMP

(dB)

10

25

0

-3

12

0

-16

20

21

-29

7

21

9

24

0

-4

11

0

-17

19

21

-30

6

21

8

23

0

-5

10

0

-18

18

21

-31

5

21

7

22

0

-6

9

0

-19

17

21

-32

4

21

6

21

0

-7

8

0

-20

16

21

-33

3

21