7. OVERPRESSURE PLUG, PART NO. 1010.

This ALL-AMERICAN Sterilizer is equipped

with an additional safety device which is the

Overpressure Plug, Part No. 1010. The

purpose of the overpressure plug is to offer

an extra margin of safety whenever the

sterilizer is used. The overpressure plug is

designed to release pressure in the range of

30 to 50 PSI.

The overpressure plug is made from silicone

and is red in color and is found on the top

surface of the sterilizer cover, located directly

to the rear of the top handle, in front of Part

No. 2050CS Excess Pressure Relief Valve.

See Figures 7 and A (page 2).

For the most efficient results and best

possible performance, it is recommended that

you replace the overpressure plug every 6

months. It should always be replaced

whenever it becomes hard or deformed.

At least every month during period of use, the

opening in the cover where the overpressure

plug fits should be checked to determine that

no foreign material, residue, or buildup of

grease is present, and the opening be cleaned

with hot soapy water (a toothbrush is helpful)

to maintain a clean opening. This

cleaning/inspection is in addition, of course,

to normal daily cleaning performed after

using the unit.

The overpressure plug can be removed for

cleaning using fingers to pull it out of its

opening from the underside of the cover.

Before you re-install the overpressure plug,

check the opening in the cover to be sure that

it is absolutely free of any foreign material or

grease/residue buildup. After cleaning,

reinsert the overpressure plug by pushing the

round top side into the opening from the

underside of the cover. When the over -

pressure plug is correctly in position, the

indented portion will be visible from the

underside of the cover. Be certain to check

after inserting plug that the round top of plug

and top lip are fully thru the opening and that

the top lip is not folded under. See Figure 7.

How Part No. 2050CS

Works

Closed

Resilient seal design prevents leakage.

Sealing efficiency increases with

increased pressure up to cracking

pressure. Metal-to-metal seat on low

pressure side supports spring load,

prevents sticking.

Open

When system pressure overcomes spring

force, poppet opens, momentarily

exposing variable orifice between poppet

and body to pass increasing flow with

minimum pressure rise without

blowdown.

Resealing

Resilient seal automatically establishes

line of contact with spherical seat. Seal

provides dead tight reseal very close to

cracking pressure.

Operating characteristics of the No.

2050CS excess pressure relief valve are:

A. Zero leakage to 95-98% of cracking

pressure.

B. Increased sealing efficiency as

pressure increases. Resilient "Q" ring

seal is forced against metal seat as

pressure increases up to set cracking

pressure.

C. Cracking pressure accuracy. Valves

are preset to required cracking

pressure of 26 PSI.

Fig. 7

Part No. 1010

Overpressure Plug

Top Lip

Cover top side

Top of Plug is rounded

Under side of plug is

indented