Cleaning with air
Blow directly on the contaminated surface with moderate pressure, clean com-
pressed air.
View a live image with the camera after blowing. If debris is still present, repeat
the process until it is determined that the particulate cannot be dislodged. If
this is the case, proceed to the contact cleaning technique.
Contact cleaning
Only use this method if the above air cleaning method does not sufficiently
clean the surface. Use 99% pure isopropyl alcohol and clean cotton swabs. Wet
the swab in the alcohol. Quickly wipe the optics in a single stroke. Prolonged
exposure of alcohol on the swab can cause the swab glue to loosen and transfer
to the optic glass. Do not reuse the same swab. Repeat this process until the
debris is removed. If this process fails to remove the debris, contact AVT.
Caution
Do not exceed 6 bar (90 psi). If using canned air, approxi-
mately ~ 4.8 bar (70 psi) when full, do not shake or tilt the
can, as extreme changes in temperature due to sudden cold
air can crack the optic glass.
- Prosilica GT4907 Prosilica GT4905 Prosilica GT3400 Prosilica GT3300 Prosilica GT2750 Prosilica GT2450 Prosilica GT2300 Prosilica GT2050 NIR Prosilica GT2050 Prosilica GT2047 Prosilica GT2000 NIR Prosilica GT2000 Prosilica GT1920 Prosilica GT1910 Prosilica GT1660 Prosilica GT1600 Prosilica GT1380 Prosilica GT1290