Dynasonics DDFXD Doppler Ultrasonic Flow Meter User Manual
Page 39

06-DPP-UM-00147 02/12
39
PIPE INSIDE DIAMETER
PIPE ID -- Pipe Inside Diameter Entry (Value)
ENGLSH
(Inches)
METRIC
(Millimeters)
Enter the pipe inside diameter in inches if INCH was selected as ID UNITS; in millimeters if MM was se-
lected.
DISPLAY MODE
DISPLAY -- Display Mode Selection (Choice)
RATE
TOTAL
BOTH
DIAG
To display only the Flow Rate, select RATE. To display only the Flow Total, select TOTAL. To alternately
display the Flow Rate and the Total, select BOTH. By selecting BOTH, the display will switch between RATE
and TOTAL every 7 seconds.
The DIAG selection places the display in the diagnostics mode. When selected, the display will show the
measured frequency,, the gain setting and the signal strength.
ENGINEERING UNITS FOR RATE
RATE UNT -- Engineering Units for Flow Rate (Choice)
Velocity in Linear Feet
(VEL FEET)
Cubic Meters
(CUBIC ME)
Velocity in Linear Meters
(VEL MTRS)
Millions of Metric Liters
(MEGLTRS)
Gallons
(GALLONS)
Acre Feet
(ACRE FT)
Liters
(LITERS)
Oil Barrels (42 Gallons)
(OIL BARR)
Millions of Gallons
(MGAL)
Liquor Barrels (31.5 Gallons (LIQ BARR)
Cubic Feet
(CUBIC FT)
Pounds
(LBS)
Millions of Cubic Feet
(M CU FT)
Kilograms
(KGS)
Select a desired engineering unit for flow rate measurements.
When Pounds (LBS) or Kilograms (KGS) is selected, the specific gravity for the fluid type must be netered
for the SP GRAV setup parameter.