Solidsense ii, Series – Brooks Instrument SolidSense II User Manual
Page 25

Installation and Operation Manual
X-PR-SolidSense II-PT-eng
Part Number 541B160AAG
July, 2011
SolidSense II
®
Series
2-9
Section 2 Installation
Position the SolidSense II pressure transducer with the electrical connector
facing pointing upward. Using a small, flat-bladed screw driver, turn zero
adjustment screw in a clockwise direction to increase the signal. Turn the
screw in a counterclockwise direction to decrease the signal.
Figure 2-9 Zero Adjustment Screw Location on the (Standard
SolidSense II pressure tranducers)
Zero Adjustment Screw
2-10 Zero Adjustment
The SolidSense II pressure transducer is factory calibrated and does not
normally need a field adjustment. If a field adjustment becomes necessary,
adjust the zero as noted below.
2-10-1 Zero Adjustment (SolidSense II Standard Pressure Transducers)
On standard SolidSense II pressure transducer models, the zero
adjustment screw is located on the body as shown in Figure 2-8.