Installation and Operation Manual
X-CM-QmB-eng
Part Number: 541B029AAG
June, 2011
QmB Series IP40, IP66, IP66XP
QS-4
3
2
Quick Start Instructions IP66XP
1
4
D-CONNECTOR PINOUTS
FUNCTION
Setpoint Common
0-5 Vdc Flow Signal Output
Alarm Output
* 4-20 mA Flow Signal Output
+14 Vdc to _27 Vdc Power Supply
Not Used
* 4-20 mA Setpoint Input (+)
0-5 Vdc Setpoint Input (+)
Power Supply Common
Signal Output Common
+5 Volt
Output
Reference
Valve Override Input
* 4-20 mA or 0-5 Vdc Density or Temp.
Not Used
Not Used
* DO NOT APPLY POWER TO THESE PINS