Burkert Type 8619 User Manual
Page 29

27
Installationandwiring
ME
RE
RE
GD
CG
FE
SE
TS
TS
1 2 3 4 5 6 7 8 9
pH measurement
electrode
Reference
electrode
Equipotential electrode
Temperature
sensor
black
(1)
translucid
(1)
blue
(2)
brown
(2)
white
(2)
black
(2)
FE = functional earth
(1)
Colour of the wires in the Bürkert connection cables with order codes 561904, 561905 or 561906.
(2)
Colour of the wires in the Pt1000 sensor with order code 427023 and its Bürkert connection cable with order code
427113.
Figure 13 :
Wiring a Bürkert 8200 type sensor and a Pt1000 temperature sensor in symmetrical mode
ME
RE
RE
GD
CG
FE
SE
TS
TS
1 2 3 4 5 6 7 8 9
Strap (not delivered)
pH measurement electrode
Reference
electrode
Equipotential
electrode
(recommended)
Temperature
sensor
black
(1)
translucid
(1)
FE = functional earth
(1)
Colour of the wires in Bürkert connection cables with order codes 561904, 561905 or 561906.
Figure 14 :
Wiring a pH sensor and a Pt100 or Pt1000 temperature sensor in asymmetrical mode to a pH/ORP module
Type 8619
english