Burkert Type 8619 User Manual

Page 95

Advertising
background image

93

Adjustmentandcommissioning

The multiCELL displays the calibration result.

1)

Rinse
7.001 pH
-59.6 mV
25.00 °C

→

Immerse the clean probe in the first
buffer solution: the multiCELL displays
the measured pH of the solution.

→

Enter the pH of the buffer solution (indicated on the bottle)

→

Wait until the measured pH stabilises.

Calibration

Mx:pH/ORP

pH manual calib.

OK

F4

1st point

7.035 pH
7.000 pH

pH Calibr. result

Zero: 7.000 pH
Slope:-59.15mV/pH

Calibrate a 2nd

point?

NO

→

Rinse the probe.

→

confirm rinsing by "OK" when the pH
reaches the desired value.

OK

F4

2nd point

2.967 pH
2.000 pH

→

Immerse the clean probe in the second buffer solution:
the multiCELL displays the measured pH of the solution.

→

Enter the pH of the second buffer solution (indicated on the
bottle)

→

Wait until the measured pH stabilises.

OK

F4

pH Calibr. result

Zero: 7.000 pH
Slope:-59.15mV/pH

YES

The multiCELL displays the
calibration result.

1)

1)

• a possible "warning" message indicates either an error in the buffer solution or the ageing of the probe.

• a possible "error" message indicates that the probe must be replaced.

Type 8619

english