Start-Up

Standard presetting:

Maximum stroke volume

Set the flow rate by turning the adjusting screw.

→

Fix the set flow rate with locking screw.

→

functions

8.3.

control diagram

8.3.1.

T0

T5

T4

T3

T2

T1

T6

ms

15

5

10

15

15

V1

V2

V3

T1

0

0

Control diagram

Fig. 6:

Via the parameters (T0 – T6), the times for the valve control are set.
By changing the interval times (T6 – T1), the frequency can be
changed.

times for maximum throughput:
While maintaining the times illustrated in the figure "Control
diagram" and an interval time of 0 ms, the maximum frequency
(15 Hz or 800 cycles/min) will be reached.
This corresponds to a throughput of approximately 4 ml/min with a
set maximum stroke volume.

recommended times for maximum accuracy:
For dosing with the highest accuracy (Tolerance and reproducibility)

≤ ± 2%), the following times are recommended:

T0

T5

T4

T3

T2

T1

T6

ms

50

50

30

50

50

T1

0

x

This corresponds to a throughput of approximately 1.15 ml/min with
a set maximum stroke volume.