Figure 5-1. troubleshooting flow chart – Fluid Components International FLT Series Rack Mount User Manual
Page 33
Advertising

FLUID COMPONENTS INTL
CHAPTER 5 - TROUBLESHOOTING
FLT
Series FlexSwitch
Rack Mount
5 - 4
Doc. No. 06EN003250 Rev. N/C
Problem
Verify Power to
Instrument
Turn Power OFF,
Remove Control
Circuit
Verfify HTR &
RTD Resist.
Install Control
Circuit, Turn Power
ON
Yellow LED
on?
Switch to CAL mode
Activate Alarm #2's LED
(green) and relay
END
Check Continuity of
the Field Wiring
STOP - Call
FCI for Repair
Repair Field
Wiring
STOP - Call
FCI for Repair
STOP - Call
FCI for Repair
STOP - Call
FCI for Repair
fail
fail
pass
fail
pass
fail
pass
pass
fail
Activate Alarm #1's
LED (red) and relay
STOP - Call
FCI for Repair
fail
pass
C00203-1
Return to RUN mode
Figure 5-1. Troubleshooting Flow Chart